[Consideration of Lens of the Eye Protection during Mammography with the Eyelid and the Face Guard]
Honami Sekizawa1, Takayuki Igarashi2, Emiko Yamamoto1
1Department of Radiology, Chiba University Hospital.
Purpose:
In 2011, the International Commissionon Radiological Protection (ICRP) recommended reducing the threshold dose for the lens. Therefore, it is important to reduce the lens exposure dose in medical exposures. In a cranio-caudal (CC) view of mammography, the patient' s lens receives scattered radiation. In this study, we investigated scatter dose around the lens during mammography and reviewed the simple and easy protection methods of the lens.
Methods:
Optically stimulated luminescence (OSL) dosimeters were placed in front of the device to obtain scattered radiation intensity distribution. The human phantom was placed in the same way as the CC positioning, and BR-12 phantoms with a thickness of 40 mm was placed on the FPD. Then, the scatter dose around the lens was measured using an OSL dosimeter. In order to confirm the change in the scatter radiation dose by the face guard (FG) and eyelid, we measured and compared under the same conditions the presence of FG and adipose tissue about 1 mm thick assuming the eyelid.
Results:
Scatter radiation intensity decreased around the FG. When the FG was installed, the scatter dose was reduced about 33%, and when the adipose tissue was pasted on the OSL dosimeter, the scatter dose was reduced about 29%.
Conclusion:
This study suggested that eye closure during mammography was effective in reducing lens exposure. In the future, we would like to expect further protective effects by increasing the thickness of FG and reviewing the materials.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
05:00Intense Pulsed Light for the Treatment of Dry Eye Owing to Meibomian Gland Dysfunction
Published on: April 1, 2019
Related Concept Videos
PPE Use in Healthcare Settings I: Donning
Accessory Structures of the Eye