Dorsa Majdpour1, Michael A Tsoukas2, Jean-François Yale2
1Department of Biomedical Engineering, McGill University, Montréal, Québec, Canada; The Research Institute of McGill University Health Centre, Montréal, Québec, Canada.
Objectives:
A fully automated insulin-pramlintide-glucagon artificial pancreas that alleviates the burden of carbohydrate counting without degrading glycemic control was iteratively enhanced until convergence through pilot experiments on adults with type 1 diabetes.
Methods:
Nine participants (age, 37±13 years; glycated hemoglobin, 7.7±0.7%) completed two 27-hour interventions: a fully automated multihormone artificial pancreas and a comparator insulin-alone artificial pancreas with carbohydrate counting. The baseline algorithm was a model-predictive controller that administered insulin and pramlintide in a fixed ratio, with boluses triggered by a glucose threshold, and administered glucagon in response to low glucose levels.
Results:
The baseline multihormone dosing algorithm resulted in noninferior time in target range (3.9 to 10.0 mmol/L) (71%) compared with the insulin-alone arm (70%) in 2 participants, with minimal glucagon delivery. The algorithm was modified to deliver insulin and pramlintide more aggressively to increase time in range and maximize the benefits of glucagon. The modified algorithm displayed a similar time in range for the multihormone arm (79%) compared with the insulin-alone arm (83%) in 2 participants, but with undesired glycemic fluctuations. Subsequently, we reduced the glucose threshold that triggers glucagon boluses. This resulted in inferior glycemic control for the multihormone arm (81% vs 91%) in 2 participants. Thereafter, a model-based meal-detection algorithm to deliver insulin and pramlintide boluses closer to mealtimes was added and glucagon was removed. The final dual-hormone system had comparable time in range (81% vs 83%) in the last 3 participants.
Conclusion:
The final version of the fully automated system that delivered insulin and pramlintide warrants a randomized controlled trial.
More Related Videos
Related Concept Videos
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are co-secreted in...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Type I Diabetes I: Introduction
Type I Diabetes II: Pathophysiology