Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Web-Based Smartphone Algorithm for Calculating Blood Pressure From Photoplethysmography Remotely in a General Adult
Paul Holyoke1, Karthika Yogaratnam1, Elizabeth Kalles1
1SE Research Centre, SE Health, Markham, ON, Canada.
Background:
Outside of a clinical setting, oscillometric devices make remote monitoring of blood pressure and virtual care more convenient and feasible. HeartBeat Technologies Ltd developed a novel approach to measuring blood pressure remotely after an initial blood pressure reading by a nurse using the conventional measurement method. Using a finger pulse oximeter, a photoplethysmogram wave is transmitted by Bluetooth to a smartphone or tablet. A smartphone app (MediBeat) transmits the photoplethysmogram to a server for analysis by a proprietary algorithm-the person's current blood pressure is sent back to the smartphone and to the individual's health care provider.
Objective:
This study sought to determine whether the HeartBeat algorithm calculates blood pressure as accurately as required by the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) for validation of blood pressure measuring devices.
Methods:
ESH-IP2 requirements, modified to conform to a more recent international consensus statement, were followed. The ESH-IP2 establishes strict guidelines for the conduct and reporting of any validation of any device to measure blood pressure, including using the standard manual blood pressure instrument as a comparator and specific required accuracy levels for low, medium, and high ranges of blood pressure readings. The consensus statement requires a greater number of study participants for each of the blood pressure ranges. The validation of the accuracy of the algorithm was conducted with a Contec CMS50EW pulse oximeter and a Samsung Galaxy XCover 4 smartphone.
Results:
The differences between the HeartBeat-calculated and the manually measured blood pressures of 62 study participants did not meet the ESH-IP2 standards for accuracy for either systolic or diastolic blood pressure measurements. There was no discernible pattern in the inaccuracies of the HeartBeat-calculated measurements.
Conclusions:
The October 4, 2019 version of the HeartBeat algorithm, implemented in combination with the MediBeat app, a pulse oximeter, and an Android smartphone, was not sufficiently accurate for use in a general adult population.
Trial Registration:
ClinicalTrials.gov NCT04082819; http://clinicaltrials.gov/ct2/show/NCT04082819.
More Related Videos
07:51Hydra, a Computer-Based Platform for Aiding Clinicians in Cardiovascular Analysis and Diagnosis
Published on: September 26, 2018
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: