Related Experiment Video
Updated: Nov 8, 2025

Monitoring Influenza Virus Survival Outside the Host Using Real-Time Cell Analysis
Published on: February 20, 2021
In-line monitoring of surfactant clearance in viral vaccine downstream processing
Jessie Payne1,2, James Cronin3, Manjit Haer1
1Analytical Sciences, Sanofi Pasteur, Toronto, Canada.
Purpose:
The goal of this study is to examine the suitability of in-line infrared measurements to monitor, in real-time, surfactant concentration in the viral vaccine drug substance during a 50KDa tangential flow filtration (TFF) process.
Methods:
A ReactIR™ 702L instrument was used to gather spectra of process off-line samples and reference materials to assess the feasibility of monitoring surfactant concentration during a TFF process in real-time. Both univariate and multivariate models were used to evaluate the off-line sample data and were found to be in good agreement with surfactant concentration values obtained by HPLC. These results were used as justification for a real-time TFF experiment with live process material.
Results:
Small scale ReactIR experiments with process material demonstrated that a multivariate model using the 1300 cm-1 to 1000 cm-1 spectral region can be used to predict surfactant concentrations between TFF exchanges 8 to 15.
Conclusion:
The results of this study demonstrated suitability of an in-line infrared measurement to monitor surfactant concentration in the viral vaccine drug substance between exchanges 8-15 of a 50 kDa tangential flow filtration process. The preliminary multivariate model used for this work can be further optimized for the in-line use at manufacturing scale.