The reliability of radiation dose display of a computed tomography scanner
V Forss1,2, H Yli-Ollila3, J Vatanen2
1Tampere University, Faculty of Medicine and Health Technology, Tampere, Finland.
Background:
Internationally, the typical allowed difference between the measured radiation dose and dose reported by a computed tomography (CT) scanner is ±20 %. The objective is to describe a method in order to analyse this difference in a CT scanner in the Emergency Department of Kanta-Häme Central Hospital, and to calculate a correction factor for more comparable radiation dose values in further studies.
Methods:
Ten intra-day radiation dose measurements were performed with undisturbed setting. Measurement reports on differences between measured and displayed dose were gathered from the vendor maintenance and supervising authority over a 12-year period. Additionally, two in-house measurements were made. A total of 18 datapoints were collected, with some differences in measurement settings. Data were also analysed against imaging parameters, ambient air pressure and time to identify trends or associations in the variation of the discrepancy.
Results:
Measured doses were generally lower than displayed doses. Differences between displayed and measured doses varied between -3.46 and -0.10 %, with a mean of -1.26 % in the intra-day measurements, and between +4.65 and -17.3 %, with a mean of -7.53 % in the long-term data. There were no trends nor connections in the variations.
Conclusion:
Since the acceptable difference between the radiation dose display and the measured dose is relevant, the average difference for every CT scanner should be determined before radiation dose studies, especially when comparing multiple scanners.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT