In vitro use of arthroscopic irrigation tubing for rapid fluid administration
1Department of Emergency Medicine, University of Louisville School of Medicine, Kentucky 40292.
The American Journal of Emergency Medicine
|July 1, 1988
Abstract:
The in vitro use of arthroscopic irrigation tubing (AIT) to achieve intravenous flow rates superior to those of previously reported studies are described. Using AIT with a peripheral type of 12 GA. catheter, flow rates of 934 cc/min with gravity and 1,928 cc/min with pressure were achieved. AIT connected to a 12 FR. central-type catheter yielded flow rates in excess of 2,000 cc and 3,000 cc/min with gravity and pressure, respectively. Such a delivery system may be lifesaving in situations where there are limited venous access sites or when there is a need for rapid volume resuscitation with limited personnel.