Dosimetric impact of rectum and bladder anatomy and intrafractional prostate motion on hypofractionated prostate
M Roch1, A Zapatero2, P Castro3
1Department of Medical Physics, Hospital La Princesa, Health Research Institute IIS-IP, Diego de León 62, 28006, Madrid, Spain. maria.roch@salud.madrid.org.
Objective:
The objective of this study was to evaluate the dosimetric impact on hypofractionated prostate radiation therapy of two geometric uncertainty sources: rectum and bladder filling and intrafractional prostate motion.
Materials And Methods:
This prospective study included 544 images (375 pre-treatment cone-beam CT [CBCT] and 169 post-treatment CBCT) from 15 prostate adenocarcinoma patients. We recalculated the dose on each pre-treatment CBCT once the positioning errors were corrected. We also recalculated two dose distributions on each post-treatment CBCT, either using or not intrafractional motion correction. A correlation analysis was performed between CBCT-based dose and rectum and bladder filling as well as intrafraction prostate displacements.
Results:
No significant differences were found between administered and planned rectal doses. However, we observed an increase in bladder dose due to a lower bladder filling in 66% of treatment fractions. These differences were reduced at the end of the fraction since the lower bladder volume was compensated by the filling during the treatment session. A statistically significant reduction in target volume coverage was observed in 27% of treatment sessions and was correlated with intrafractional prostate motion in sagittal plane > 4 mm.
Conclusions:
A better control of bladder filling is recommended to minimize the number of fractions in which the bladder volume is lower than planned. Fiducial mark tracking with a displacement threshold of 5 mm in any direction is recommended to ensure that the prescribed dose criteria are met.