Related Experiment Video
Updated: Nov 7, 2025

07:34
Author Spotlight: Advancing Prostate Cancer Research Through Improved Tissue Sampling and Biobanking
Published on: November 17, 2023
913
Assessment of Agreement between Two Difference Prostate-Specific Antigen Assay Modalities
Jae Hoon Chung1, Minsu Park2, Hyun Cho3
1Samsung Medical Center, Department of Urology, Sungkyunkwan University School of Medicine, Seoul 06351, Korea.
Biology
|April 30, 2021
Summary
Prostate-specific antigen (PSA) assay modalities show excellent overall agreement but poor agreement in specific low and high ranges. Differences in PSA values between modalities may be clinically significant for prostate cancer detection.
Area of Science:
- Urology
- Oncology
- Biomarker Analysis
Background:
- Prostate-specific antigen (PSA) is a key biomarker for prostate cancer (PCa).
- Controversy exists regarding PSA's utility due to variability between assay modalities.
- Standardized validation of PSA assay methods is crucial for accurate PCa diagnosis.
Purpose of the Study:
- To evaluate the agreement between two PSA assay modalities.
- To identify specific PSA concentration ranges where agreement is poor.
- To determine the clinical relevance of observed differences in PSA measurements.
Main Methods:
- Analysis of PSA levels from 4810 patients using two distinct assay modalities.
- Statistical evaluation using intra-class correlation coefficient (ICC) and weighted kappa analysis.
- Linear regression to assess the association and predict values between modalities.
Main Results:
- Excellent overall agreement between PSA assay modalities (ICC: 0.999, kappa: 0.951).
- Poor agreement observed in specific PSA sub-groups: 0.05-0.1, 0.15-0.2, 1.5-2.0, and 2.0-2.5 ng/mL.
- Linear regression indicated significant differences, e.g., 0.2 ng/mL in modality B corresponded to 0.258 ng/mL in modality A.
Conclusions:
- While overall agreement is excellent, significant discrepancies exist in critical PSA ranges.
- The observed differences between PSA assay modalities can be clinically meaningful.
- Further validation and consideration of assay-specific values are necessary for reliable PCa biomarker interpretation.