Related Experiment Video
Updated: Nov 7, 2025

Clinical Assessment of Spatiotemporal Gait Parameters in Patients and Older Adults
Published on: November 7, 2014
Accuracy and Repeatability of Spatiotemporal Gait Parameters Measured with an Inertial Measurement Unit
Jorge Posada-Ordax1, Julia Cosin-Matamoros2, Marta Elena Losa-Iglesias1
1Facultad de Ciencias de la Salud, Universidad Rey Juan Carlos, 28933 Madrid, Spain.
Abstract:
In recent years, interest in finding alternatives for the evaluation of mobility has increased. Inertial measurement units (IMUs) stand out for their portability, size, and low price. The objective of this study was to examine the accuracy and repeatability of a commercially available IMU under controlled conditions in healthy subjects. A total of 36 subjects, including 17 males and 19 females were analyzed with a Wiva Science IMU in a corridor test while walking for 10 m and in a threadmill at 1.6 km/h, 2.4 km/h, 3.2 km/h, 4 km/h, and 4.8 km/h for one minute. We found no difference when we compared the variables at 4 km/h and 4.8 km/h. However, we found greater differences and errors at 1.6 km/h, 2.4 km/h and 3.2 km/h, and the latter one (1.6 km/h) generated more error. The main conclusion is that the Wiva Science IMU is reliable at high speeds but loses reliability at low speeds.