Related Experiment Video
Updated: Oct 19, 2026

Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
Comparison of three methods for determination of N-nitrosopyrrolidine in fried dry-cured and pump-cured bacon
A J Malanoski1, W J Smith, T Phillipo
1U.S. Department of Agriculture, Food Safety and Inspection Service, Washington, DC 20250.
Abstract:
Three methods, the mineral oil distillation (MOD), the dry column (DC), and a low temperature vacuum distillation (LTVD), for the determination of N-nitrosopyrrolidine in dry-cured and pump-cured bacon were compared. Each method uses the thermal energy analyzer for the determinative step. The coefficients of variation for repeatability were 10.3% (MOD), 7.2% (DC), and 9.1% (LTVD) for the dry-cured bacon study and 8.7% (MOD), 8.5% (DC), and 7.1% (LTVD) for the pump-cured bacon study. The pooled coefficients of variation for between-method reproducibility were 11.8% for the dry-cured bacon and 10.8% for the pump-cured bacon. The pooled coefficients of variation for repeatability were 9.0% for the dry-cured bacon and 8.2% for the pump-cured bacon. These values compare favorably with the values from previous collaborative or validation studies of the individual methods, and the methods can be considered to be equivalent.
More Related Videos
07:38A General Method for Detecting Nitrosamide Formation in the In Vitro Metabolism of Nitrosamines by Cytochrome P450s
Published on: September 25, 2017
08:23Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022