Related Experiment Video
Updated: Nov 7, 2025

Author Spotlight: Optimizing EAS with Long Electrodes for Enhanced Cochlear Coverage and Hearing Preservation
Published on: October 11, 2024
Adult Users of the Oticon Medical Neuro Cochlear Implant System Benefit from Beamforming in the High Frequencies
Bianca Bastos Cordeiro1, Marcos Roberto Banhara2, Carlos Maurício Cardeal Mendes1
1Institute of Health Sciences, Federal University of Bahia, Salvador 40110-100, Brazil.
Abstract:
The Oticon Medical Neuro cochlear implant system includes the modes Opti Omni and Speech Omni, the latter providing beamforming (i.e., directional selectivity) in the high frequencies. Two studies compared sentence identification scores of adult cochlear implant users with Opti Omni and Speech Omni. In Study 1, a double-blind longitudinal crossover study, 12 new users trialed Opti Omni or Speech Omni (random allocation) for three months, and their sentence identification in quiet and noise (+10 dB signal-to-noise ratio) with the trialed mode were measured. The same procedure was repeated for the second mode. In Study 2, a single-blind study, 11 experienced users performed a speech identification task in quiet and at relative signal-to-noise ratios ranging from -3 to +18 dB with Opti Omni and Speech Omni. The Study 1 scores in quiet and in noise were significantly better with Speech Omni than with Opti Omni. Study 2 scores were significantly better with Speech Omni than with Opti Omni at +6 and +9 dB signal-to-noise ratios. Beamforming in the high frequencies, as implemented in Speech Omni, leads to improved speech identification in medium levels of background noise, where cochlear implant users spend most of their day.