Related Experiment Videos
Effects of gating modes on the analysis of left ventricular function in the presence of heart rate variation
1Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor.
Abstract:
Variations in cardiac cycle length during ECG-gated radionuclide ventriculography cause errors in measurement of left ventricular (LV) diastolic function. We used a computer model to investigate the relative accuracy of forward gating (FG), backward gating (BG), and three methods of combined forward-backward gating (FBG). LV time-activity curves (TAC) were simulated with total cycle lengths randomly varied (1 s.d. = +/- 10%). To match the known differential effects of cycle-length variation on various parts of the cardiac cycle, 90% this variation was placed in diastasis and 10% in the rapid filling phase. Three hundred cycles of TACs were summed to form one "study". A total of 25 studies were performed and systolic and diastolic parameters were determined for each. These were compared to those of the mean length TAC. With a 20% cycle-length acceptance window, error in peak filling rate by FG was 0.7% root-mean-squared (RMS) compared to 18.5% RMS by BG. With a 5% window, RMS error by FG was 0.2% and 2.4% RMS by BG. Combined FBG algorithms gave intermediate results. We conclude that the simplest and most accurate method of analysis is to acquire forward gated TACs with appropriate cycle-length windows.