Impact of combining data from multiple instruments on performance of patient-based real-time quality control
Qianqian Zhou1, Tze Ping Loh2, Tony Badrick3
1Engineering Cluster, Singapore Institute of Technology, Singapore, Singapore.
Introduction:
It is unclear what is the best strategy for applying patient-based real-time quality control (PBRTQC) algorithm in the presence of multiple instruments. This simulation study compared the error detection capability of applying PBRTQC algorithms for instruments individually and in combination using serum sodium as an example.
Materials And Methods:
Four sets of random serum sodium measurements were generated with differing means and standard deviations to represent four simulated instruments. Moving median with winsorization was selected as the PBRTQC algorithm. The PBRTQC parameters (block size and control limits) were optimized and applied to the four simulated laboratory data sets individually and in combination.
Results:
When the PBRTQC algorithm were individually optimized and applied to the data of the individual simulated instruments, it was able to detect bias several folds faster than when they were combined. Similarly, the individually applied algorithms had perfect error detection rates across different magnitudes of bias, whereas the error detection rates of the algorithm applied on the combined data missed smaller biases. The performance of the individually applied PBRTQC algorithm performed more consistently among the simulated instruments compared to when the data were combined.
Discussion:
While combining data from different instruments can increase the data stream and hence, increase the speed of error detection, it may widen the control limits and compromising the probability of error detection. The presence of multiple instruments in the data stream may dilute the effect of the error when it only affects a selected instrument.
More Related Videos
06:32Author Spotlight: Automated Deep Brain Stimulation for Parkinson's Disease - Exploring the Possibilities and Challenges of Home Monitoring
Published on: July 14, 2023
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Errors occurring during blood pressure monitoring
Several factors...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Introduction to Statistical Process Control
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...