Single-shot RARE with Dixon: Application to robust abdominal imaging with uniform fat and water separation at 3T
Xinzeng Wang1,2, Durga Udayakumar1,3, Yin Xi1,4
1Department of Radiology, UT Southwestern Medical Center, Dallas, Texas, USA.
Purpose:
To develop a true single shot turbo spin echo (SShTSE) acquisition with Dixon for robust T2 -weighted abdominal imaging with uniform fat and water separation at 3T.
Methods:
The in-phase (IP) and out-of-phase (OP) echoes for Dixon processing were acquired in the same repetition time of a SShTSE using partial echoes. A phase-preserved bi-directional homodyne reconstruction was developed to compensate the partial echo and the partial phase encoding of SShTSE. With IRB approval, the SShTSE-Dixon was compared against standard SShTSE, without and with fat suppression using spectral adiabatic inversion recovery (SPAIR) in 5 healthy volunteers and 5 patients. The SNR and contrast ratio (CR) of spleen to liver were compared among different acquisitions.
Results:
The bi-directional homodyne reconstruction successfully minimized ringing artifacts because of partial acquisitions. SShTSE-Dixon achieved uniform fat suppression compared to SShTSE-SPAIR with fat suppression failures of 1/10 versus 10/10 in the axial plane and 0/5 versus 5/5 in the coronal plane, respectively. The SNRs of the liver (12.2 ± 4.9 vs. 11.7 ± 5.2; P = .76) and spleen (25.9 ± 11.6 vs. 23.7 ± 9.7; P = .14) were equivalent between fat-suppressed images (SShTSE-Dixon water-only and SShTSE-SPAIR). The SNRs of liver (14.4 ± 5.7 vs. 13.4 ± 5.0; P = .60) and spleen (26.5 ± 10.1 vs. 25.7 ± 8.5; P = .56) were equivalent between non-fat-suppressed images (SShTSE-Dixon IP and SShTSE). The CRs of spleen to liver were also similar between fat-suppressed images (2.6 ± 0.4 vs. 2.5 ± 0.5; P =.92) and non-fat-suppressed images (2.3 ± 0.6 vs. 2.2 ± 0.4; P =.84).
Conclusion:
SShTSE-Dixon generates robust abdominal T2 -weighted images at 3T with and without uniform fat suppression, along with perfectly co-registered fat-only images in a single acquisition.
More Related Videos
Related Concept Videos
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...