Quantitative evaluation of microvacuole formation in five intraocular lens models made of different hydrophobic
Timur M Yildirim1, Sonja K Schickhardt1, Qiang Wang1,2
1The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany.
Abstract:
In this laboratory study, we assessed the resistance to microvacuole (glistening) formation in hydrophobic intraocular lenses (IOLs). Glistenings were induced in five lenses each of five different hydrophobic acrylic IOL models, using an established in vitro laboratory model: 800C (Rayner, Worthing, UK), AcrySof SN60WF (Alcon, Fort Worth, USA), Tecnis ZCB00 (Johnson & Johnson Vision, Santa Ana, USA), Vivinex XY1 (Hoya, Tokyo, Japan) and CT Lucia 611P (Zeiss, Oberkochen, Germany). We evaluated the number of microvacuoles per square millimeter (MV/mm2) in the central part of each IOL. Results were analyzed statistically, and mean glistening numbers were ranked, with the highest in the SN60WF which had 66.0 (±45.5) MVs/mm, followed by the 611P with 30.7 (±8.4) MVs/mm2. The 800C and XY1 showed comparable values of 2.0 (±3.6) and 2.7 (±2.4) MVs/mm2, respectively. ZCB00 had the lowest number with 0.9 (±0.6) MVs/mm2. This study shows that the resistance to glistening formation differs depending on the hydrophobic acrylic copolymer composition of the IOL material. Some IOLs from current clinical use are still prone to develop glistenings whereas others, including the ZCB00, 800C and XY1 show high resistance to microvacuole formation.
More Related Videos
Related Concept Videos
In Vitro Drug Dissolution: Compendial Testing Models II
In Vitro Drug Dissolution: Compendial Testing Models I