Oxygen delivery failure due to improper installation of carbon dioxide absorbent canister: a case report
Sher-Lu Pai1, Christopher B Robards1, Kevin T Riutort1
1Mayo Clinic, Department of Anesthesiology and Perioperative Medicine, Jacksonville, Florida, United States.
Abstract:
The CLIC system in the Dräger Apollo anesthesia workstation allows a successful pre-use machine checkout without the presence of a carbon dioxide absorbent canister. It also allows the canister to be changed without interrupting controlled ventilation. However, this canister can be easily installed improperly with the CLIC adapter. We report a case in which a patient could not be ventilated by mask after the induction of general anesthesia, resulting in oxygen desaturation before successful ventilation was achieved with a bag valve mask. This case illustrates the importance of a leak test after components of the breathing circuit are changed.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: