Ultra-low-dose lung multidetector computed tomography in children - Approaching 0.2 millisievert
Sebastian Tschauner1, Michael Zellner2, Sarah Pistorius2
1University Children's Hospital Zurich, Department of Imaging, Zurich, Switzerland; Medical University of Graz, Department of Radiology, Division of Pediatric Radiology, Graz, Austria.
Purpose:
To compare objective and subjective parameters in image quality and radiation dose of two MDCTs (helical 64 detector CT vs. axial 256 detector CT) in paediatric lung CT.
Methods:
Radiation dose and image quality were compared between non-enhanced lung CT from a helical 64-slice multidetector CT (MDCT 1) and a 256-slice scanner (MDCT 2) with axial wide-cone acquisition and using deep learning image reconstruction. In 23 size-matched paediatric studies (age 2-18 years) from each scanner, the radiation exposure, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), image sharpness and delineation of small airways were assessed. Subjective image quality was rated by 6 paediatric radiologists.
Results:
While MDCT 2 provided higher SNR and CNR, subjective image quality was not significantly different between studies from both scanners. Radiation exposure was lower in studies from MDCT 2 (CTDIvol 0.26 ± 0.14 mGy, effective dose 0.23 ± 0.11 mSv) than from MDCT 1 (CTDIvol 0.96 ± 0.52 mGy, effective dose 1.13 ± 0.58 mSv), p < 0.001. Despite lower radiation dose for the scout images, the relative scout-scan-ratio increased from 2.64 ± 1.42 % in MDCT 1 to 6.60 ± 5.03 % in MDCT 2 (p = 0.001).
Conclusions:
By using latest scanner technology effective radiation dose can be reduced to 0.1-0.3 mSv for lung CT in children without compromising image quality. Scout image dose increasingly accounts for substantial portions of the total scan dose and needs to be optimized. In children CT should be performed on state-of-the-art MDCT scanners with size-adapted exposure protocols and iterative reconstruction.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...