Comparison of image quality and lesion diagnosis in abdominopelvic unenhanced CT between reduced-dose CT using deep
Xiao Wang1, Fuling Zheng1, Ran Xiao1
1From the Department of Radiology, State Key Laboratory of Complex Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, China.
Purpose:
To compare image quality and lesion diagnosis between reduced-dose abdominopelvic unenhanced computed tomography (CT) using deep learning (DL) post-processing and standard-dose CT using iterative reconstruction (IR).
Method:
Totally 251 patients underwent two consecutive abdominopelvic unenhanced CT scans of the same range, including standard and reduced doses, respectively. In group A, standard-dose data were reconstructed by (blend 30 %) IR. In group B, reduced-dose data were reconstructed by filtered back projection reconstruction to obtain group B1 images, and post-processed using the DL algorithm (NeuAI denosing, Neusoft medical, Shenyang, China) with 50 % and 100 % weights to obtain group B2 and B3 images, respectively. Then, CT values of the liver, the second lumbar vertebral centrum, the erector spinae and abdominal subcutaneous fat were measured. CT values, noise levels, signal-to-noise ratios (SNRs), contrast-to-noise ratios (CNRs), radiation doses and subjective scores of image quality were compared. Subjective evaluations of low-density liver lesions were compared by diagnostic results from enhanced CT or Magnetic Resonance Imaging.
Results:
Groups B3 and B1 showed the lowest and highest noise levels, respectively (P < 0.001). The SNR and CNR in group B3 were highest (P < 0.001). The radiation dose in group B was reduced by 71.5 % on average compared to group A. Subjective scores in groups A and B2 were highest (P < 0.001). Diagnostic sensitivity and confidence for liver metastases in groups A and B2 were highest (P < 0.001).
Conclusions:
Reduced-dose abdominopelvic unenhanced CT combined with DL post-processing could ensure image quality and satisfy diagnostic needs.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...