Evaluation of SARS-CoV-2 total antibody detection via a lateral flow nanoparticle fluorescence immunoassay
Mamdouh Sibai1, Daniel Solis1, Katharina Röltgen1
1Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA.
Background:
The coronavirus disease 2019 (COVID-19) endgame may benefit from simple, accurate antibody testing to characterize seroprevalence and immunization coverage.
Objectives:
To evaluate the performance of the lateral flow QIAreach anti-SARS-CoV-2 Total rapid nanoparticle fluorescence immunoassay compared to reference isotype-specific IgG, IgM, and IgA SARS-CoV-2 ELISA using S1 or receptor binding domain (RBD) as antigens.
Study Design:
A diagnostic comparison study was carried out using 154 well-characterized heparin plasma samples. Agreement between assays was assessed by overall, positive, and negative percent agreement and Cohen's kappa coefficient.
Results:
Overall agreement between the QIAreach anti-SARS-CoV-2 Total and any anti-spike domain (S1 or RBD) antibody isotype was 96.0 % (95 % CI 89.8-98.8), the positive percent agreement was 97.6 % (95 % CI 91.0-99.9), the negative percent agreement was 88.2 % (95 % CI 64.4-98.0). The kappa coefficient was 0.86 (95 % CI 0.72 to 0.99).
Conclusion:
The QIAreach anti-SARS-CoV-2 Total rapid antibody test provides comparable performance to high-complexity, laboratory-based ELISA.