Hemodynamic study of unenhanced magnetic resonance angiography using spatial labeling with multiple inversion pulses
Xiao Chen1, Xiaoyan Meng1, Di Zhu1
1Department of Radiology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
Quantitative Imaging in Medicine and Surgery
|May 3, 2021
Summary
This study shows that displayed vascular length linearly correlates with blood suppression inversion time (BSP TI) and flow rate in MRA phantoms. This establishes a basis for quantitatively assessing blood flow using unenhanced MRA techniques.
Area of Science:
- Medical Imaging
- Biophysics
- Cardiovascular Imaging
Background:
- Unenhanced Magnetic Resonance Angiography (MRA) is crucial for vascular assessment.
- Spatial Labeling with Multiple Inversion Pulses Sequence (SLEEK) is an MRA technique.
- Understanding the relationship between MRA parameters and hemodynamics is essential for quantitative analysis.
Purpose of the Study:
- To explore the functional relationship between displayed vascular length, blood suppression inversion time (BSP TI), and flow velocity in a phantom.
- To provide a theoretical basis for quantitatively assessing vascular hemodynamic responses using unenhanced MRA and SLEEK.
Main Methods:
- A phantom model with a polyethylene catheter and saline flow was used.
- Flow rates ranged from 0.0 to 2.4 mL/s.
- SLEEK was performed with varying BSP TI values (50-1800 ms) and flow rates, followed by regression analysis.
Main Results:
- A significant linear relationship was observed between displayed vascular length and BSP TI for flow rates from 0.0 to 2.0 mL/s (R²=0.754-0.988).
- A significant linear relationship was also found between displayed vascular length and flow rate for BSP TI values from 50 to 1100 ms (R²=0.834-0.996).
- No significant linear relationship was detected between displayed vascular length and flow rate at BSP TI of 1200 ms.
Conclusions:
- Displayed vascular length demonstrates a linear relationship with BSP TI within a specific flow range (0.0-2.0 mL/s).
- Displayed vascular length is linearly related to flow rate across a range of BSP TI values (50-1100 ms).
- Flow rate can be quantitatively assessed based on displayed vascular length in MRA.
Keywords:
Unenhanced MR angiographyblood flowblood suppression inversion timemultiple inversion pulsesphantomvascular lengthMore Related Videos
Related Concept Videos
Magnetic Resonance Imaging
8.3K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
8.3K
Imaging Studies for Cardiovascular System IV: CMRI
172
Cardiovascular magnetic resonance imaging, or CMRI, is a non-invasive diagnostic test that employs a magnetic field and radiofrequency waves to create precise images of the heart and arteries. It provides comprehensive information about cardiac anatomy, function, perfusion, and tissue characterization without ionizing radiation.IndicationsCMRI diagnoses various heart conditions, including tissue damage from heart attacks, ischemic heart disease, myocarditis, aortic issues (tears, aneurysms,...
172