Organic contamination detection for isotopic analysis of water by laser spectroscopy
Cody Millar1, Kim Janzen1, Magali F Nehemy1
1Global Institute for Water Security, School of Environment and Sustainability, University of Saskatchewan, 11 Innovation Boulevard, Saskatoon, SK, S7N 3H5, Canada.
Rationale:
Hydrogen and oxygen stable isotope ratios (δ2 H, δ17 O, and δ18 O values) are commonly used tracers of water. These ratios can be measured by isotope ratio infrared spectroscopy (IRIS). However, IRIS approaches are prone to errors induced by organic compounds present in plant, soil, and natural water samples. A novel approach using 17 O-excess values has shown promise for flagging spectrally contaminated plant samples during IRIS analysis. A systematic assessment of this flagging system is needed to prove it useful.
Methods:
Errors induced by methanol and ethanol water mixtures on measured IRIS and isotope ratio mass spectrometry (IRMS) results were evaluated. For IRIS analyses both liquid- and vapour-mode (via direct vapour equilibration) methods are used. The δ2 H, δ17 O, and δ18 O values were measured and compared with known reference values to determine the errors induced by methanol and ethanol contamination. In addition, the 17 O-excess contamination detection approach was tested. This is a post-processing detection tool for both liquid and vapour IRIS triple-isotope analyses, utilizing calculated 17 O-excess values to flag contaminated samples.
Results:
Organic contamination induced significant errors in IRIS results, not seen in IRMS results. Methanol caused larger errors than ethanol. Results from vapour-IRIS analyses had larger errors than those from liquid-IRIS analyses. The 17 O-excess approach identified methanol driven error in liquid- and vapour-mode IRIS samples at levels where isotope results became unacceptably erroneous. For ethanol contaminated samples, a mix of erroneous and correct flagging occurred with the 17 O-excess method. Our results indicate that methanol is the more problematic contaminant for data corruption. The 17 O-excess method was therefore useful for data quality control.
Conclusions:
Organic contamination caused significant errors in IRIS stable isotope results. These errors were larger during vapour analyses than during liquid IRIS analyses, and larger for methanol than ethanol contamination. The 17 O-excess method is highly sensitive for detecting narrowband (methanol) contamination error in vapour and liquid analysis modes in IRIS.
More Related Videos
13:38Laser-Induced Fluorescence Emission L.I.F.E. as Novel Non-Invasive Tool for In-Situ Measurements of Biomarkers in Cryospheric Habitats
Published on: October 26, 2019
09:42Fabrication of a Dipole-assisted Solid Phase Extraction Microchip for Trace Metal Analysis in Water Samples
Published on: August 7, 2016
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
High-Performance Liquid Chromatography: Types of Detectors
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...