Multiacquisition Variable-Resonance Image Combination Selective Can Improve Image Quality and Reproducibility for
Yuta Fujiwara1, Tomoaki Sasaki2, Yuki Muto2
1Division of Clinical Radiology Service, Okayama Central Hospital.
Abstract:
The aim of this study is to evaluate how metallic artifacts in the lumbar spine can affect images obtained from magnetic resonance (MR) sequences. We performed a phantom experiment by scanning an agar containing an orthopedic metallic implant using 64-channel multidetector row computed tomography (CT) and a 3-tesla MR unit. We compared the reproducibility in each measurement, enlargement or reduction ratio of the CT and MR measurements, and signal deviation in each voxel from the control. The reproducibility on CT and multiacquisition variable-resonance image combination selective (MAVRIC SL) was good, but that on the other MR sequences showed either fixed bias or proportional bias. The reduction ratios of the distance between the nails were significantly smaller in MAVRIC SL than in the other MR sequences after CT measurements (p<0.001, respectively). MAVRIC SL was able to reduce the metallic artifact, permitting observation of the tissue surrounding the metal with good reproducibility.
More Related Videos
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
18:40Combined In vivo Optical and µCT Imaging to Monitor Infection, Inflammation, and Bone Anatomy in an Orthopaedic Implant Infection in Mice
Published on: October 16, 2014
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging