Related Experiment Video
Updated: Aug 13, 2026

Preparation and Testing of Impedance-based Fluidic Biochips with RTgill-W1 Cells for Rapid Evaluation of Drinking Water Samples for Toxicity
Published on: March 7, 2016
An ultrasensitive fluorimetric sensor for pre-screening of water microbial contamination risk
João Simões1, Zhaochu Yang2, Tao Dong1
1Department of Microsystems-IMS, Faculty of Technology, Natural Sciences and Maritime Sciences, University of South-Eastern Norway-USN, P.O. Box 235, Kongsberg 3603, Norway.
Abstract:
In recent years, global efforts have been directed towards the development of water safety routines, consequently demanding cost-effective sensors capable of detecting outbreaks at early stages. This work reports the development and study of an original in-field tryptophan fluorimetric sensor as a potential indicator of real-time microbial contamination in water. The sensor excitation and emission wavelengths were selected with respect to the coliform bacteria tryptophan peak; 280 nm for excitation and 330 nm for emission. The in-lab tests with standard samples show a detection limit of 4.89 nM (≈0.1 μg/l) for L-tryptophan. The sensor exhibited good linearity over three orders of magnitude and considerable detection reproducibility, which was confirmed during calibration tests. Small-scale in situ tests showed that the sensor was better correlated with coliform bacteria than other online sensors such as turbidity. This suggests that the fluorimetric tryptophan sensor can be integrated into early warning systems that quickly assess changes in water microbial quality.