Attitude is everything: keep probe pitch neutral during side-fire prostate biopsy. A simulator study
Zhou Zhang1,2, Samsun Lampotang1,3,4,5, Yichao Yu6
1Centre for Safety, Simulation and Advanced Learning Technologies, University of Florida College of Medicine, Gainesville, FL, USA.
BJU International
|May 7, 2021
Summary
A new simulated prostate biopsy technique significantly reduced core deviation, aiming to decrease false-negative results and improve early cancer detection. This learnable method enhances accuracy in systematic prostate biopsies.
Area of Science:
- Urology
- Medical Simulation
- Oncology
Background:
- Systematic prostate biopsy (sPBx) aims to reduce false-negative rates for clinically significant prostate cancer.
- Freehand techniques can lead to core deviation from the intended template schema.
- Reducing deviation is crucial for accurate cancer diagnosis and timely treatment.
Purpose of the Study:
- To develop and validate a learnable technique to minimize core deviation during freehand, side-fire systematic prostate biopsy (sPBx).
- To reduce prostate biopsy (PBx) false-negatives by improving core sampling accuracy.
- To facilitate earlier diagnosis and treatment of prostate cancer.
Main Methods:
- A pitch-neutral, side-fire, transrectal ultrasonography (TRUS)-guided sPBx technique was devised using a simulator with 3D visualization.
- Thirty-four trainees practiced the technique in a left lateral decubitus position.
- Deviation from the template and TRUS probe pitch were recorded; mastery was defined as template deviation ≤5.0 mm.
Main Results:
- Mean template deviation decreased from 8.0 ± 2.7 mm at baseline to 4.5 ± 0.6 mm after mastering the technique (P < 0.001).
- Mean absolute pitch decreased from 8.2 ± 4.1° to 4.5 ± 2.7° (P = 0.001).
- Mastery was achieved after practicing approximately 4 sets, with no significant increase in procedure time.
Conclusions:
- A pitch-neutral side-fire technique effectively reduces template deviation in simulated freehand TRUS-guided sPBx.
- This technique is teachable and learnable, with potential to reduce PBx false-negatives in clinical settings.
- The technique is already being integrated into urology residency training programs.