Predictive Intelligent Control of Oxygenation (PRICO) in preterm infants on high flow nasal cannula support: a
Koen P Dijkman1, Thilo Mohns2, Jeanne P Dieleman3
1Neonatology, Maxima Medical Centre, Veldhoven, The Netherlands k.dijkman@mmc.nl.
Objective:
To investigate the efficacy of automated control of inspired oxygen (FiO2) by Predictive Intelligent Control of Oxygenation (PRICO) on the Fabian ventilator in maintaining oxygen saturation (SpO2) in preterm infants on high flow nasal cannula (HFNC) support.
Design:
Single-centre randomised two-period crossover study.
Setting:
Tertiary neonatal intensive care unit.
Patients:
27 preterm infants (gestational age (GA) <30 weeks) on HFNC support with FiO2 >0.25.
Intervention:
A 24-hour period on automated FiO2-control with PRICO compared with a 24-hour period on routine manual control (RMC) to maintain a SpO2 level within target range of 88%-95% measured at 30 s intervals.
Main Outcome Measures:
Primary outcome: time spent within target range (88%-95%).
Secondary Outcomes:
time spent above and below target range, in severe hypoxia (SpO2 <80%) and hyperoxia (SpO2 >98%), mean SpO2 and FiO2 and manual FiO2 adjustments.
Results:
15 patients received PRICO-RMC and 12 RMC-PRICO. The mean time within the target range increased with PRICO: 10.8% (95% CI 7.6 to 13.9). There was a decrease in time below target range: 7.6% (95% CI 4.2 to 11.0), above target range: 3.1% (95% CI 2.9 to 6.2) and in severe hypoxia: 0.9% (95% CI 1.5 to 0.2). We found no difference in time spent in severe hyperoxia. Mean FiO2 was higher during PRICO: 0.019 (95% CI 0.006 to 0.030). With PRICO there was a reduction of manual adjustments: 9/24 hours (95% CI 6 to 12).
Conclusion:
In preterm infants on HFNC support, automated FiO2-control by PRICO is superior to RMC in maintaining SpO2 within target range. Further validation studies with a higher sample frequency and different ventilation modes are needed.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: