Rupture of an epidural filter connector during bolus administration of local anesthetic: a case report
Related Experiment Video
Updated: Nov 5, 2025

A Mobile Outside-in Technique of Transforaminal Lumbar Endoscopy for Lumbar Disc Herniations
Published on: August 7, 2018
Daniel A Nahrwold1, Aaron R Muncey2, Nasrin N Aldawoodi2
1H. Lee Moffitt Cancer Center & Research Institute, University of South Florida Morsani College of Medicine, 12902 USF Magnolia Dr, Tampa, FL, 33612, USA. daniel.nahrwold@moffitt.org.
Background:
Epidural catheters are routinely placed for many surgical procedures and to treat various pain conditions. Known complications arising from epidural catheter equipment malfunction include epidural pump failure, epidural catheter shearing, epidural catheter connector failure, epidural filter connector cracking, and loss-of-resistance syringe malfunction. Practitioners need to be aware of these potentially dangerous complications and take measures to mitigate the chances of causing significant patient harm. We report on the complete breakage of an epidural filter connector during epidural bolus administration of local anesthetic by hand with a syringe.
Case Presentation:
A B. Braun Perifix® epidural catheter was placed in a 73-year-old male scheduled for radical prostatectomy. During the operation, a continuous infusion of local anesthetic was administered through the epidural catheter in addition to general endotracheal anesthesia. At the conclusion of surgery and after extubation, the patient endorsed incisional pain. The epidural filter connector broke in half as a bolus of local anesthetic was administered by hand with a syringe. The local anesthetic sprayed widely throughout the room as the fragmented epidural filter connector became a projectile object that recoiled and struck the patient.
Conclusions:
This incident placed the patient and surrounding healthcare providers at substantial risk for injury and infection from the fractured epidural filter connector becoming a projectile object and from the local anesthetic spray. The most plausible cause of this event was from a large amount of pressure being applied to the filter connector. This may have occurred by excessive force being applied by hand to the syringe, by the presence of a clogged filter, or by the catheter being kinked or blocked proximal to the filter. Being aware of this deleterious complication and potentially modifying existing epidural bolus techniques, such as using smaller syringes with less applied force and checking all epidural components vigilantly prior to and during bolus administration, can help anesthesia providers deliver the safest possible care to patients with epidural catheters.
More Related Videos
06:59Intrathecal Application of a Fluorescent Dye for the Identification of Cerebrospinal Fluid Leaks in Cochlear Malformation
Published on: February 29, 2020
04:33Spinal Hernia Repair and Cauda Equina Repositioning After Lumbar Decompression under Three-Dimensional Microscopy: A Case Report and Literature Review
Published on: November 8, 2024
Related Concept Videos
Local Anesthetics: Clinical Application as Epidural Anesthesia
Since epidural anesthetics can be infused through an epidural catheter, all types of drugs, including short-acting ones, can be administered. Chloroprocaine and lidocaine are examples of short and long-duration anesthetics, respectively. Bupivacaine...
Local Anesthetics: Clinical Application as Spinal Anesthesia
Local Anesthetics: Clinical Application as Intravenous Regional Anesthesia
One of the advantages of...
Local Anesthetics: Adverse Effects
Once absorbed into the systemic circulation, local anesthetics can affect the organs that depend on the functioning of sodium...
Local Anesthetics: Pharmacokinetics
Local Anesthetics: Clinical Application as Surface, Infiltration, and Conduction Block Anesthesia