In-plane flow velocity quantification along the phase encoding axis in MRI.

J L Duerk1, P M Pattany

  • 1Cleveland Metropolitan General Hospital, Radiology Dept., Ohio 44109.

Magnetic Resonance Imaging
|May 1, 1988
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Control of intravascular catheters using an array of active steering coils.

Medical physics·2011
Same author

Magnetic resonance imaging-guided renal artery stent placement in a Swine model: comparison of two tracking techniques.

Acta radiologica (Stockholm, Sweden : 1987)·2008
Same author

Quantitative analysis of the effects of physiologic brain motion on point-resolved spectroscopy.

AJNR. American journal of neuroradiology·2006
Same author

Modified block uniform resampling (BURS) algorithm using truncated singular value decomposition: fast accurate gridding with noise and artifact reduction.

Magnetic resonance in medicine·2001
Same author

A method for fast 3D tracking using tuned fiducial markers and a limited projection reconstruction FISP (LPR-FISP) sequence.

Journal of magnetic resonance imaging : JMRI·2001
Same author

Segmented k-space and real-time cardiac cine MR imaging with radial trajectories.

Radiology·2001

This study introduces a novel MRI technique for precise in-plane flow quantification. The motion artifact suppression technique (MAST) enables accurate assessment of blood flow velocity and volume, improving MR angiography.

Area of Science:

  • Medical Imaging
  • Biomedical Engineering
  • Fluid Dynamics

Background:

  • In-plane flow quantification in MRI is crucial for assessing vessel patency, volume flow rate, and flow velocity.
  • Current techniques have limitations in accuracy and velocity range for MR angiography.

Purpose of the Study:

  • To develop and validate a novel MRI method for in-plane flow quantification.
  • To improve accuracy and expand velocity measurement limits compared to existing methods.

Main Methods:

  • Utilized motion artifact suppression technique (MAST) gradients for spin refocusing.
  • Encoded velocity along the phase encoding axis, minimizing artifacts and higher-order effects.
  • Conducted flow phantom studies with varying flow rates (0.6-2.5 GPM) and velocities (13.2-55.3 cm/sec).

Related Experiment Videos

Main Results:

  • Achieved accurate flow rate measurements with errors ≤8% across the studied range.
  • Demonstrated elimination or reduction of flow voids due to spin dephasing.
  • Successfully measured velocities within a wider range (13.2-55.3 cm/sec) and Reynolds numbers (2,700-11,500).

Conclusions:

  • The MAST-based MRI technique provides accurate in-plane flow quantification.
  • This method enhances MR angiography capabilities by overcoming limitations of previous techniques.
  • Offers potential for improved clinical assessment of vascular conditions.