Related Experiment Video
Updated: Nov 5, 2025

How to Build a Laser Speckle Contrast Imaging LSCI System to Monitor Blood Flow
Published on: November 11, 2010
Quantitative laser speckle auto-inverse covariance imaging for robust estimation of blood flow
Abstract:
We present a quantitative model to provide robust estimation of the decorrelation time using laser speckle auto-inverse covariance. It has the advantages of independence from the statistical sample size, speckle size, static scattering, and detector noise. We have shown cerebral blood flow imaging through an intact mouse skull using this model. Phantom experiments and two animal models, middle cerebral artery occlusion, and cortical spreading depression were used to evaluate its performance.
More Related Videos
07:20A Novel Approach to Overcome Movement Artifact When Using a Laser Speckle Contrast Imaging System for Alternating Speeds of Blood Microcirculation
Published on: August 30, 2017
08:00Meso-Scale Particle Image Velocimetry Studies of Neurovascular Flows In Vitro
Published on: December 3, 2018