Related Experiment Video
Updated: Nov 5, 2025

A Simplified Stepwise Approach to Echo Guidance during Percutaneous Mitral Valve Repair
Published on: October 16, 2021
Early cusp tear of the second-generation Trifecta valve with Glide Technology
Akimasa Morisaki1, Yosuke Takahashi1, Shinsuke Nishimura1
1Department of Cardiovascular Surgery, Osaka City University Graduate School of Medicine, Osaka, Japan.
Abstract:
We herein report a case of early cusp tear of the second-generation Trifecta valve with Glide Technology. A 69-year-old woman underwent aortic valve replacement with a Trifecta valve with Glide Technology and partial arch replacement for severe aortic valve stenosis caused by a bicuspid valve and ascending aortic and aortic arch aneurysms 2 years previously. During follow-up, she suddenly developed dyspnoea and a heart murmur. Echocardiography revealed severe aortic valve regurgitation caused by structural valve deterioration. We performed re-aortic valve replacement with a mechanical valve through re-median sternotomy. Intraoperative examination revealed a bottom tear of the left coronary cusp without infection or pannus formation.
Related Concept Videos
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Stenosis III: Medical Management
Mitral Regurgitation I: Introduction
Mitral Regurgitation III: Medical Management
Mitral Valve Prolapse III: Nursing Management