Delivery of Inhaled Nitric Oxide During MRI to Ventilated Neonates and Infants
Bradley G Carter1, Rachel Swain2, Jaime Hislop2
1Clinical Technology Service, Neonatal and Paediatric Intensive Care Units, Royal Children's Hospital, Parkville, Victoria, Australia. bradley.carter@rch.org.au.
Background:
Many pediatric and neonatal ICU patients receive nitric oxide (NO), with some also requiring magnetic resonance imaging (MRI) scans. MRI-compatible NO delivery devices are not always available. We describe and bench test a method of delivering NO during MRI using standard equipment in which a NO delivery device was positioned in the MRI control room with the NO blender component connected to oxygen and set to 80 ppm and delivering flow via 12 m of tubing to a MRI-compatible ventilator, set up inside the MRI scanner magnet room.
Methods:
For our bench test, the ventilator was set up normally and connected to an infant test lung to simulate several patients of differing weight (ie, 4 kg, 10 kg, 20 kg). The NO blender delivered flows of 2-10 L/min to the ventilator to achieve a range of NO and oxygen concentrations monitored via extended tubing. The measured values were compared to calculated values.
Results:
A range of NO concentrations (12-41 ppm) and F values (0.67-0.97) were achieved during the bench testing. The additional flow increased delivered peak inspiratory pressure and PEEP by 1-5 cm H2O. Calculated values were within acceptable ranges and were used to create a lookup table.
Conclusions:
In clinical use, this system can safely generate a range of NO flows of 15-42 ppm with an accompanying F range of 0.34-0.98.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...