Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Tension band plates have greater risks of complications in temporary epiphysiodesis
You-Hung Cheng1, Wei-Chun Lee1, Yi-Feng Tsai1
1Department of Pediatric Orthopedics, Chang Gung Memorial Hospital, Chang Gung University, Taoyuan, Taiwan.
Purpose:
This study aimed to compare the efficacy of decreasing leg-length discrepancy (LLD) and postoperative complications between tension band plates (TBP) and percutaneous transphyseal screws (PETS).
Methods:
This retrospective study reviewed LLD patients who underwent temporary epiphysiodesis at the distal femur and/or proximal tibia from 2010 to 2017 (minimum two years follow-up). Efficacy of decreasing LLD was assessed one and two years postoperatively. Complications were classified with the modified Clavien-Dindo-Sink complication classification system. Knee deformities were assessed by percentile and zone of mechanical axis across the tibial plateau.
Results:
In total, 53 patients (25 boys, 28 girls) underwent temporary epiphysiodesis (mean age, 11.4 years). The efficacy of decreasing LLD at two years between the TBP (n = 38) and PETS (n = 15) groups was comparable. Seven grade III complications were recorded in six TBP patients and in one PETS patient who underwent revision surgeries for knee deformities and physis impingement. Four grade I and two grade II complications occurred in the TBP group. The mechanical axis of the leg shifted laterally in the PETS group and medially in the TBP groups (+7.1 percentile versus -4.2 percentile; p < 0.05). Shifting of the mechanical axis by two zones was noted medially in four TBP patients and laterally in two PETS patients.
Conclusion:
More implant-related complications and revision surgeries for angular deformities were associated with TBP. A tendency of varus and valgus deformity after epiphysiodesis using TBP and PETS was observed, respectively. Patients and families should be informed of the risks and regular postoperative follow-up is recommended.
Level Of Evidence:
Level III.