Related Experiment Video
Updated: Nov 4, 2025

Experimental Procedure for Warm Spinning of Cast Aluminum Components
Published on: February 1, 2017
Aluminium release and fluid warming: provocational setting and devices at risk
Thorsten Perl1, N Kunze-Szikszay2, A Bräuer2
1Department of General, Visceral and Pediatric Surgery, University Medical Center Goettingen, Robert-Koch-Str. 40, D- 37075, Goettingen, Germany. tperl@gwdg.de.
Background:
Fluid warming, recommended for fluid rates of > 500 ml h-1, is an integral part of patient temperature management strategies. Fluid warming devices using an uncoated aluminium containing heating element have been reported to liberate aluminium resulting in critical aluminium concentrations in heated fluids. We investigated saline solution (0.9%), artificially spiked with organic acids to determine the influence of fluid composition on aluminium release using the uncoated enFlow® device. Additionally, the Level1® as a high volume fluid warming device and the ThermoSens® device were investigated with artificial spiked fluid at high risk for aluminum release and a clinically used crystalloid solution.
Results:
Saline solution spiked with lactate more than acetate, especially at a non neutral pH, led to high aluminium release. Next to the enFlow® device, aluminium release was observed for the Level1® device, but not for the coated ThermoSens®-device.
Conclusion:
Uncoated aluminium containing fluid warming devices lead to potentially toxic levels of aluminium in heated fluids, especially in fluids with non-neutral pH containing organic acids and their salts like balanced electrolyte solutions.
Related Concept Videos
Factors Affecting Body Temperature
Factors may include:
Decreased Body Temperature
Increased Body Temperature
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Fluid Pressure
According to Pascal's law, a fluid at rest will generate equal pressure in all directions. This pressure is measured as a force per unit area, and its magnitude depends on the fluid's specific...