Evaluation of the relationship between phantom position and computed tomography dose index in cone beam computed
Hiroyuki Ueno1,2, Kosuke Matsubara2, Akihiro Takemura2
1Department of Radiology, Takaoka City Hospital, Takaoka, Toyama, Japan.
Abstract:
This study aims to investigate the influence of the phantom position on weighted computed tomography dose index (CTDIw ) in cone beam computed tomography (CBCT) when assuming breast irradiation. Computed tomography dose index (CTDI) was measured by the x-ray volume imaging of CBCT using parameters for image-guided radiation therapy (IGRT) in right breast irradiation. The measurement points of CTDI ranged from 0 (center) to 16 cm in the right-left (RL) direction, and from 0 (center) to 7.5 cm in the anterior-posterior (AP) direction, which assumed right breast irradiation. A nonuniform change exists in the relative value of CTDIw when the phantom deviated from the isocenter of CBCT. The CTDIw was ~30% lower compared with the value at the isocenter of CBCT when the phantom deviated 7.5 and 16 cm at the AP and RL directions, respectively. This study confirmed the influence of the phantom position on the CTDI values of CBCT. The CTDI measured at the isocenter of CBCT overestimates that measured at the irradiation center of the breast.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography