A rare case of persistent pseudohypobicarbonatemia arising from chemistry analyzer-specific interference
Lucy Ma1, Zhen Zhao1, Sabrina E Racine-Brzostek1
1Department of Pathology and Laboratory Medicine, Weill Cornell Medicine, New York, NY, USA; New York-Presbyterian Hospital, Weill Cornell Medicine, New York, NY, USA.
Background:
Major discrepancies between concentrations of serum total carbon dioxide (tCO2) obtained from chemistry analyzers and calculated bicarbonate from blood gas analyzers should prompt laboratory investigation. Here, we present a rare case of pseudohypobicarbonatemia unrelated to the common causes such as hypertriglyceridemia and hyperproteinemia, but was caused by a low concentration of paraproteins.
Case:
A 75-year-old man with persistent fevers was found to have a low concentration of serum tCO2 (<10 mmol/l) with a normal pH and calculated bicarbonate concentrations (23.5 mmol/l) from the blood gas analyzer. His serum tCO2 concentrations remained critically low throughout hospitalization without any evidence of acidosis. Serum tCO2 levels were measured via Siemens ADVIA Chemistry XPT system.
Results:
Mixing studies revealed non-linearity of serum tCO2, suggesting the presence of interfering substances. Triglyceride concentrations were normal. Serum electrophoresis revealed a 0.4 mg/dl M-protein. The patient's serum tCO2 concentrations were repeated on different chemistry analyzer platforms - including Siemens, Roche, and Abbott - which demonstrated that the interference was specific to the Siemens chemistry analyzer. Serum tCO2 was significantly elevated after ultrafiltration of paraprotein, which confirmed the root cause of pseudohypobicarbonatemia.
Conclusion:
Laboratory professionals should be aware that spuriously low serum tCO2 concentrations may result from unique interfering substances, such as paraproteins, that are both patient- and chemistry analyzer-specific.
More Related Videos
09:34Improved Polymerase Chain Reaction-restriction Fragment Length Polymorphism Genotyping of Toxic Pufferfish by Liquid Chromatography/Mass Spectrometry
Published on: September 20, 2016
08:59Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Disorders of Acid-Base Balance
Respiratory Acidosis and Alkalosis
Respiratory acidosis occurs due to an increase in the partial pressure of carbon dioxide PCO2 in the blood. It often arises from shallow breathing or impaired gas exchange caused by...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Emission Spectroscopy: Interference