Related Experiment Video
Updated: Nov 4, 2025

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Three-year follow-up after less-invasive left ventricular assist device exchange to HeartMate 3™
Jasmin S Hanke1, Silvia Mariani1, Ali S Merzah1
1Department of Cardiac, Thoracic, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
Background:
Device exchange to a newer generation left ventricular assist device (LVAD) offers the opportunity to benefit from improved adverse events profiles. We present the three-year results of a patient cohort undergoing VAD upgrades to a new generation device focusing on outcomes and adverse events.
Methods:
We present the first series of patients who underwent LVAD upgrade to HeartMate 3™ (Abbott Laboratories; Abbott Park, IL, USA). All operations were performed less invasively. Follow-up time was three years after LVAD exchange.
Results:
Overall four HeartMate II™ (Abbott Laboratories; Abbott Park, IL, USA) and two HVAD patients underwent LVAD upgrade. In five cases severe infection of the VAD led to device exchange (83%, 5/6). Three-year survival after LVAD exchange was 100% (6/6). In the follow-up examinations one patient showed a single syncope and several low flow alarms (1/6). The remaining five patients showed no technical malfunctions of the LVAD or hemodynamic adverse events (5/6). Four out of five patients whose devices had to be changed due to an infection suffered a local re-infection (4/5), which did not require any further surgical intervention. Four patients were successfully transplanted, and two patients were still on device support at three years after LVAD exchange.
Conclusions:
Three-year outcomes and adverse events after LVAD exchange to HeartMate 3™ (Abbott Laboratories). show excellent results. The superior hemocompatibility in terms of pump thrombosis makes the HM3 a favored choice in case of LVAD exchange due previous pump thrombosis. However, in cases of exchange due to device infection the risk of re-infection remains high.