Water-cooled microwave ablation array for bloodless rapid transection of the liver
Qiang Yang1, Wei Li1, Zubing Chen2
1Department of Pancreatic Surgery, Renmin Hospital of Wuhan University, Wuhan University, Wuhan, China.
Background:
Microwaves (MWs) deliver relatively high temperatures into biological tissue and cover a large ablation zone. This study aims to evaluate the efficacy and effectiveness of water-cooled double-needle MW ablation arrays in assisting the hepatic transection of an in vivo pig model.
Methods:
Our research program comprised computer modeling, tissue-mimicking phantom experiments, and in vivo pig liver experiments. Computer modeling was based on the finite element method (FEM) to evaluate ablation temperature distributions. In tissue-mimicking phantom and in vivo pig liver ablation experiments, the performances of the water-cooled MW ablation array and conventional clamp crushing liver resection were compared.
Results:
FEM showed that the maximum lateral ablation diameter at 100 W output and a duration of 60 s was 3 cm (assessed at 50 °C isotherm). In the phantom, the maximum transverse ablation diameter of the double-needle MW ablation increased rapidly to 3 cm in 60 s at 50 W. The blood loss and blood loss per transection area in Group A were significantly lower than those in Group B (18 (7-26) ml vs. 34 (19-57) ml, and 2.4 (2-3.1) ml/cm2 vs. 6.9 (3.2-8.3) ml/cm2, respectively) (p < 0.05). The transection speed in Group A (2.6(1.9-3.8) cm2/min) was significantly faster than that in Group B (1.7(1.1-2.2) cm2/min) (p < 0.05).
Conclusion:
In this experimental model, the new water-cooled MW array-assisted liver resection (LR) has the potential advantage of less blood loss and rapid removal than the conventional LR.