Determination of a Pu-Be source neutron spectrum at Czech Metrology Institute
1Czech Metrology Institute, Okružní 31, 638 00, Brno, Czech Republic.
Abstract:
Contrary to the radionuclide neutron sources of 252Cf and 241Am-Be type, which are most widely used for the dosimetric instrumentation calibrations, the spectral source strength and spectrum averaged fluence to dose equivalent conversion coefficients of Pu-Be source are not recommended by an international standard. This work describes the determination of those parameters for the Pu-Be source used at Czech Metrology Institute (CMI) by means of a Bonner spectrometer in combination with neutron emission rate measurement in a manganese bath.