High Accuracy Heartbeat Detection from CW-Doppler Radar Using Singular Value Decomposition and Matched Filter.
Yuki Iwata1, Han Trong Thanh2, Guanghao Sun1
1Graduate School of Informatics and Engineering, The University of Electro-Communications (UEC), Tokyo 182-8585, Japan.
Sensors (Basel, Switzerland)
|June 2, 2021
Summary
A new matched filter method accurately extracts heart rate from radar signals, overcoming challenges posed by body motion. This non-contact vital signs monitoring technique shows promise for home healthcare applications.
Area of Science:
- Biomedical Engineering
- Signal Processing
- Remote Sensing
Background:
- Continuous wave Doppler radar sensors (CW-DRS) offer non-contact vital signs monitoring, crucial for home healthcare.
- Extracting cardiac signals from CW-DRS data is challenging due to superimposed respiratory and body motion signals.
- Existing methods struggle to isolate precise cardiac information from complex radar measurements.
Purpose of the Study:
- To develop a novel signal processing method for accurate heart rate extraction using CW-DRS.
- To address the challenge of separating cardiac signals from superimposed motion artifacts.
- To validate the proposed method across different subject postures.
Main Methods:
- A novel matched filter approach was developed.
- The method involves adaptive template generation using singular value decomposition (SVD) of a trajectory matrix.
- Signal reconstruction is achieved through convolution of the generated template with measurement signals.
Main Results:
- The proposed method demonstrated effectiveness in extracting heart rate from CW-DRS signals.
- For supine subjects, absolute errors in heart rate and standard deviation of heartbeat interval were 1.93±1.76 bpm and 57.0±28.1 s, respectively.
- For seated subjects, errors were 9.72±7.86 bpm and 81.3±24.3 s, indicating good performance across postures.
Conclusions:
- The novel matched filter method successfully extracts cardiac information from CW-DRS signals, even with superimposed motion.
- This technique enhances the feasibility of non-contact vital signs monitoring in home healthcare settings.
- The validated method offers improved accuracy for heart rate and heartbeat interval measurements in various subject positions.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
1.9K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.9K
Doppler Effect - II
3.8K
The Doppler effect has several practical, real-world applications. For instance, meteorologists use Doppler radars to interpret weather events based on the Doppler effect. Typically, a transmitter emits radio waves at a specific frequency toward the sky from a weather station. The radio waves bounce off the clouds and precipitation and travel back to the weather station. The radio frequency of the waves reflected back to the station appears to decrease if the clouds or precipitation are moving...
3.8K
Equipments Used To Measure Blood Pressure
2.7K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.7K
Pulse rhythm
1.0K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.0K