A Radiolucent Electromagnetic Tracking System for Use with Intraoperative X-ray Imaging
Kilian O'Donoghue1, Herman Alexander Jaeger1, Padraig Cantillon-Murphy1,2
1Tyndall National Institute, Dyke Parade, T12 R5CP Cork, Ireland.
Sensors (Basel, Switzerland)
|June 2, 2021
Summary
This study introduces a new radiolucent electromagnetic navigation system for surgical procedures. The novel design significantly reduces X-ray artifacts, improving surgical imaging while maintaining accuracy.
Area of Science:
- Medical Imaging
- Surgical Navigation
- Biomedical Engineering
Background:
- Electromagnetic tracking (EM) is crucial for minimally invasive surgery, often used with X-ray technology.
- Conventional EM systems cause X-ray artifacts and attenuation due to radiopaque materials.
- Existing systems limit the clarity of intraoperative imaging.
Purpose of the Study:
- To develop a novel radiolucent electromagnetic navigation system.
- To minimize the impact of EM tracking on X-ray imaging.
- To improve the integration of EM tracking and X-ray diagnostics in surgery.
Main Methods:
- Developed a new electromagnetic navigation system using radiolucent materials.
- Replaced traditional dense materials (copper, steel) in EM coils with low-density aluminum.
- Continued development of the Anser open-source electromagnetic tracking system.
Main Results:
- Achieved at least a 60% reduction in X-ray attenuation compared to previous designs.
- Demonstrated significantly reduced impact on X-ray images.
- Maintained system accuracy with approximately 1.5 mm RMS error.
Conclusions:
- The novel radiolucent electromagnetic navigation system effectively reduces X-ray artifacts.
- This advancement enhances the utility of EM tracking in conjunction with X-ray imaging for surgical navigation.
- The improved system offers better image clarity without compromising tracking precision.