Heated Humidifiers for Noninvasive Respiratory Support and the Risk of Burns in Neonates: A Bench Evaluation
Sebastien Fau1, Olivier Baud2, Peter Rimensberger2
1Service of Neonatology and Pediatric Intensive Care, Department of Pediatrics, University Hospital of Geneva, Geneva, Switzerland. sebastien.fau@hcuge.ch.
Background:
User errors in managing heated humidifiers (HHs) have been suggested to be a source of nasal burns in newborns treated with nasal CPAP. This study evaluated the risk of burns by reproducing 3 typical errors concerning the use of HHs.
Methods:
Six HHs were tested on a bench in a traditional nasal CPAP setup: PMH5000, Aircon (Wilamed); MR730, MR850, MR950 (Fisher & Paykel); and H900 (Hamilton). Temperature was measured at the end of the inspiratory tubing limb. Errors tested were (1) misconnection of the HH thermal probes (NoProbe), (2) absence of gas flow while the HH is on (NoFlow), and (3) unsuitable repeated acknowledgment of the HH alarm (NoAlarm). These errors were combined in 3 standardized scenarios: (1) NoProbe + NoFlow + NoAlarm; (2) NoProbe + NoAlarm, and (3) NoFlow + NoAlarm. The NoProbe + NoFlow + NoAlarm and NoProbe + NoAlarm scenarios were not tested in the H900 and MR950 because the proprietary circuits of these HHs are equipped with embedded probes.
Results:
For each HH, the highest inspiratory gas temperature (HIGT) and the rating on a self-designed risk-of-burn scale (ie, no risk, moderate risk, or severe risk) were reported. In the NoProbe + NoFlow + NoAlarm scenario, the risk was severe for the MR730, PMH5000, MR850, and Aircon, with HIGTs of > 65°C, 58°C, 56°C, and > 65°C, respectively. In the NoProbe + NoAlarm scenario, the risk was also severe for the same 4 HHs, with HIGTs of 56°C, 47°C, 56°C, and 48°C, respectively. In the NoFlow + NoAlarm scenario, the risk was severe for the PMH5000, Aircon, and H900, with HIGTs of 52°C, > 65°C, and 49°C, respectively, and moderate for the MR730, MR850, and MR950, with HIGTs of 45°C, 47°C, and 44°C, respectively).
Conclusions:
In case of misuse, 5 of the 6 tested devices presented a severe risk of inducing skin burns, whereas the MR950 presented a moderate risk.
More Related Videos
04:21A Minimally Invasive Method for Intratracheal Instillation of Drugs in Neonatal Rodents to Treat Lung Disease
Published on: August 4, 2021
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Decreased Body Temperature
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: