Agreement Between Transcutaneous Monitoring and Arterial Blood Gases During COPD Exacerbation
Kasper M Sørensen1,2, Rebecca V Leicht1,2, Christian J Carlsson1,2
1Department of Anaesthesia and Intensive Care, Bispebjerg and Frederiksberg Hospital, University of Copenhagen, Copenhagen, Denmark.
Background:
Transcutaneous measurements of CO2 and O2 ([Formula: see text], [Formula: see text]) are noninvasive and allow for continuous monitoring in adults with exacerbation of COPD, but substantial accuracy issues may exist. We investigated agreement between results of arterial blood gas analysis and transcutaneous measurements of CO2 and O2 in patients with COPD.
Methods:
Adult subjects were monitored after acute admission to a respiratory intermediate care unit or ICU due to exacerbation of COPD and with ongoing noninvasive ventilation or immediately following extubation. Monitored variables were continuous transcutaneous measurement and simultaneous routine arterial blood gas analysis. Agreement between measurements was assessed by calculating bias with 95% limits of agreement for single-point estimates of [Formula: see text] versus [Formula: see text] and versus [Formula: see text], and for changes in transcutaneous measurements between 2 time points ([Formula: see text] and [Formula: see text]). We considered limits of agreement within ± 7.5 mm Hg to be acceptable.
Results:
A total of 57 transcutaneous measurements were made in 20 subjects for comparison with concurrent arterial blood gas analysis at 36 time points. The bias (limits of agreement) for [Formula: see text] and [Formula: see text] was 2.5 mm Hg (-10.6 to 15.6 mm Hg) and 11.2 mm Hg (-28.2 to 50.6 mm Hg), respectively. The bias for [Formula: see text] and [Formula: see text] was 2.3 mm Hg (-3.8 to 8.3 mm Hg) and -5.3 mm Hg (-37.5 to 27 mm Hg), respectively.
Conclusions:
[Formula: see text] and [Formula: see text] did not accurately reflect results from arterial blood gas analyses in this study of mostly hypercapnic subjects. Agreement between changes in CO2 during the monitoring period was acceptable, however, and transcutaneous monitoring may be used for continuous monitoring of [Formula: see text] in conjunction with arterial blood gas analysis for reference.
More Related Videos
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Acute Respiratory Failure-III
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...