Related Experiment Video
Updated: Nov 3, 2025

Author Spotlight: Development of a Minimally Invasive Large-Animal Model for Reliable and Reproducible Cardiovascular Research
Published on: October 20, 2023
A Translational 'Humanised' Porcine Model for Trans-Septal Mitral Valve Interventions: the Right Pulmonary Vein
Avi Avner1, Itai Sabbag2, Adi Butnaru3
1Knowledge Farm Diagnostic Imaging, PO Box 136, 30900, Zichron Yaakov, Israel. aviavner7@gmail.com.
Abstract:
There is an important need to develop a transcatheter/trans-septal-like model which will allow deliverability of mitral valve devices at the right angulation and height above the mitral annulus similar to the human patient. A ministernotomy allowing for TEE probe to be introduced transpericardially was performed in seven animals. In three, a standard trans-septal approach was used to attempt to deliver a device to the level of the mitral plane with a human delivery system. In four cases, a Dacron tube was connected to the RPV and the device introduced to the LA. Under TEE equivalent imaging, the device was then successfully steered to the mitral plane in all cases, having failed in all trans-septal attempts. A combination of transpericardial echocardiographic imaging (TPE) and RPV approach to the LA facilitated successful human-like mitral valve device implantation.
More Related Videos
06:56Author Spotlight: Advancing Cardiac Procedure Testing Prior to Embarking on Large Animal Studies
Published on: August 25, 2023
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022