Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Effect of cement space on marginal discrepancy and retention of CAD/CAM crown
Laila Azwa Hassan1, Chui Ling Goo1
1Department of Restorative Dentistry, Faculty of Dentistry, The National University of Malaysia.
Abstract:
This research aimed to evaluate the effect of cement space on the marginal discrepancy and retention of computer-aided design/computer-aided manufacturing (CAD/CAM) crowns. A total of 30 premolar Frasaco teeth were machined to receive crowns with cement spaces of 70, 90, and 110 μm. The marginal discrepancy measurements were done before and after cementation. Pull-off test was conducted using universal testing machine (UTM). Data was analyzed using two-way mixed ANOVA with post-hoc Bonferroni test and Kruskal-Wallis test. The crowns with cement space of 70 μm showed a significantly higher absolute marginal discrepancy than those with 90 and 110 μm. No significant effect on the crown retention was found. Within the limitations of this study, modifying cement space to 90 μm and 110 μm may improve the marginal adaptation of CAD/CAM crown, whereas adjusting cement space from 70 to 110 μm did not significantly affect the crown retention.
Related Concept Videos
Soundness of Cement
Porosity in Cement Paste
The balance of water to cement in the mix is...
Aggregate Cement Ratio
Total Voids in Concrete
Fineness of Cement
Direct...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...