Incorrect Holding Angle of Dry Powder Inhaler during the Drug-Loading Step Significantly Decreases Output Efficiency
Kana Suenaga1, Daiki Hira1,2, Eri Ishido1
1College of Pharmaceutical Sciences, Ritsumeikan University.
Abstract:
It is well known that correct use of inhalers plays a critical role in optimal inhalation therapy, but the impact of incorrect inhaler use on pulmonary drug delivery has not been quantitatively evaluated. The aim of this study was to investigate the frequency of holding inhalers at incorrect angles during the drug-loading step while using Turbuhaler® and to quantify the influence of the inhaler angle on in vitro pulmonary delivery. Thirty patients prescribed Turbuhaler® at Shiga University of Medical Science Hospital were enrolled. During inhalation, the participants' inhalation techniques were assessed by clinical pharmacists. Additionally, the influence of the inhaler angle on pulmonary delivery of budesonide via Symbicort® Turbuhaler® was investigated using a Twin-Stage Liquid Impinger. Output efficiency (OE), stage 2 deposition (St2), and OE × St2 were calculated. An incorrect angle during the drug-loading step was observed in 33.3% of the participants. In vitro testing demonstrated that OE, an index of the loaded dose, significantly decreased by 73.3% at an incorrect angle, while St2, an index of the deagglomerating efficiency, was stable independent of the holding angle. OE × St2, indicating the bronchial and pulmonary drug delivery amount, decreased by 76.9%. An incorrect holding angle reduced the loaded dose, resulting in decreased pulmonary delivery. Error in the inhaler angle occurs frequently and demonstrates a considerable impact on pulmonary drug delivery. Hence, it is necessary to assess the Turbuhaler® angle during inhalation.
More Related Videos
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
04:59Author Spotlight: Developing a Disposable Dosator for Preclinical Testing of Dry Powder Inhalers in Small Animal Models
Published on: August 18, 2023
Related Concept Videos
Inhaled Medications
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Factors Influencing Drug Absorption: Pharmaceutical Parameters
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Formulation and Manufacturing Process: Physical Attributes of Generic Tablets and Capsules
Pharmaceutical Alternatives: Polymorphic Form-Related and Particle Size-Related Therapeutic Nonequivalence