Comparison of Roche and Lumipulse quantitative SARS-CoV-2 antigen test performance using automated systems for the
Yosuke Hirotsu1, Hiroki Sugiura2, Makoto Maejima3
1Genome Analysis Center, Yamanashi Central Hospital, 1-1-1 Fujimi, Kofu, Yamanashi, Japan.
Background:
Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) continues to spread worldwide. Here, we evaluated the performance of two quantitative antigen (Ag) tests, the Roche and Lumipulse Ag tests, using automated platforms.
Methods:
We collected 637 nasopharyngeal swab samples from 274 individuals. Samples were subjected to quantitative reverse transcription PCR (RT-qPCR), the Roche Ag test and Lumipulse Ag test.
Results:
When RT-qPCR was used as a reference, the overall concordance rate of the Roche Ag test was 77.1% (491/637) with 70.0% (341/487) sensitivity and 100% specificity (150/150). When inconclusive results of the Lumipulse Ag test were excluded, the overall concordance rate of the Lumipulse Ag test was 88.3% (467/529) with 84.8% (330/389) sensitivity and 97.9% (137/140) specificity. The overall concordance rate between the Roche and Lumipulse Ag tests was 97.9% (518/529) with 96.7% (322/333) sensitivity and 100% (196/196) specificity. Quantitative Ag levels determined using the Roche and Lumipulse Ag tests were highly correlated (R2 = 0.922). The Roche and Lumipulse Ag tests showed high concordance up to nine days after symptom onset, with progressively lower concordance after that.
Conclusions:
The Roche and Lumipulse Ag tests showed equivalent assay performance and represent promising approaches for diagnosing coronavirus disease 2019.