Impact of Multileaf Collimator Width on Dose Distribution in HyperArc Fractionated Stereotactic Irradiation for
Shingo Ohira1,2, Yoshihiro Ueda3, Naoyuki Kanayama3
1Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan; oohira-si@mc.pref.osaka.jp.
Background/Aim:
To assess the impact of the width of multileaf collimator (MLC) on dose distributions on HyperArc fractionated stereotactic irradiation for multiple (5-10) brain metastases.
Patients And Methods:
Twenty-one HyperArc (HA) plans were generated using the high definition (HD) MLC (2.5 mm) to deliver 30-35 Gy in 3-5 fractions (HA-HD). The HyperArc plans using Millennium (ML) MLC (5 mm) were retrospectively generated (HA-ML) using the same planning parameters with HA-HD. Dosimetric parameters between the planning target volume (PTV) and organs at risk (OARs) were compared.
Results:
The conformity index was significantly higher (p<0.0001) in the HA-HD plans (0.95±0.04) than that in the HA-ML plans (0.92±0.06). The HA-HD provided significantly lower (p<0.0001) gradient index (5.6±2.5) than HA-ML (6.2±3.5). For the brainstem and retina (right), a statistically significant difference (p<0.05) was observed between the HA-HD (12.8±10.9 and 2.8±1.7 Gy, for brainstem and retina, respectively) and HA-ML (13.6±11.1 and 3.0±1.8 Gy) plans. For the brain tissue, the HA-HD plans statistically significantly reduced dosimetric parameters (p<0.0001) in all evaluated dose range (V6Gy-V28Gy).
Conclusion:
The narrower MLC provided significantly higher conformity, steeper dose gradient, and better normal tissue sparing.