Measurement of Bioelectric Current with a Vibrating Probe
Published on: January 4, 2011
Examining electrometer performance checks with direct-current generator in a clinic: Assessment of generated charges
Naoki Kinoshita1, Hiroshi Oguchi2, Morihito Shimizu3
1Radiological Center, University of Fukui Hospital, Fukui, Japan.
Purpose:
Medical physicists use a suitable detector connected to an electrometer to measure radiotherapy beams. Each detector and electrometer has a lifetime (due to physical deterioration of detector components and electrical characteristic deterioration in electronic electrometer components), long-term stability [according to IEC 60731:2011, ≤0.5% (reference-class dosimeter)], and calibration frequency [according to Muir et al. (J Appl Clin Med Phys. 2017; 18:182-190), generally 2 years]; thus, physicists should check the electrometer and detector separately. However, to the best of our knowledge, only one study (Blad et al., Phys Med Biol. 1998; 43:2385-2391) has reported checking the electrometer independently from the detector. The present study conducts performance checks on electrometers separately from the detector in clinical settings, using an electrometer equipped with a direct current (DC) generator (EMF 521R) capable of injecting DC (effective range: ±20 pA to ±20 nA) into itself or another electrometer.
Methods:
First, to check the nonlinearity of the generated currents from ±20 pA to ±20 nA, charges generated from the DC generator were measured with the EMF 521R electrometer. Next, six reference-class electrometers classified according to IEC 60731:2011 were checked for repeatability at a current of ±20 pA or a minimum effective indicated value meeting IEC 60731:2011, as well as for nonlinearity within the current range from ±20 pA to ±20 nA.
Results:
The nonlinearities for the measured currents were less than ±0.05%. The repeatability for the six electrometers was < 0.1%. While the nonlinearity of one electrometer reached up to 0.22% at a current of -20 pA, all six electrometers displayed nonlinearities of less than ±0.1% at currents of ±100 pA or higher.
Conclusions:
This work suggests that it is possible to check the nonlinearity and repeatability of clinical electrometers with DCs above the ±30 pA level using a DC generator in a clinic.
Related Concept Videos
Voltmeter
An ideal voltmeter would have infinite resistance, so connecting it between two points in a circuit would not alter any of the currents. Real voltmeters always have...
DC Generator
Galvanometer
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform...
Potentiometer
Suppose the emf of a battery needs to be measured. If the battery is directly connected to a standard voltmeter, the measured quantity is...
Controlled-Current Coulometry: Overview
Coulometry: Overview