Comparison of the analytical performance of the PATHFAST high sensitivity cardiac troponin I using fresh whole blood
Vinajak Gopi1, Barbara Milles1, Eberhard Spanuth2
1Department of Cardiology, Angiology and Pulmonology, Heidelberg University Hospital, Heidelberg, Germany.
Objectives:
The PATHFAST hs-cTnI (high-sensitivity cardiac troponin) assay is the first point-of-care assay with a high-sensitivity designation that received FDA approval for diagnosis of myocardial infarction (MI). Testing from whole blood does not need centrifugation and therefore is faster and more convenient in the emergency room instead of plasma. However, there is sparse evidence whether point-of-care testing of Tn from whole blood is as reliable as from plasma samples.
Methods:
We investigated the agreement between plasma and whole blood hs-cTnI by using the PATHFAST hs-cTnI assay. Hs-cTnT measured on Cobas 602 in the central laboratory and compared to a final diagnosis of NSTEMI using serial hs-cTnT served as reference. We assessed biases, limits of agreement (±1.96 SD) and coefficients of correlation, and tested the discriminatory ability of the baseline sample of plasma and whole blood hs-cTnI and plasma hs-cTnT to discriminate non-ST-segment elevation myocardial infarction (NSTEMI).
Results:
A total of 224 paired fresh samples were collected simultaneously from 191 patients presenting with suspected acute coronary syndrome. There was an excellent correlation between plasma and whole blood hs-cTnI (r=0.99), and a very good inter-rater agreement (k=0.93) between elevated and normal plasma and whole blood results. Precision evaluation according to CLSI ep 15 revealed comparable coefficients of variation (CV) in whole blood and plasma. The discriminatory ability of baseline hs-cTnT, plasma and whole blood hs-cTnI was excellent (AUC 0.967, AUC 0.954 and AUC 0.953) without significant difference.
Conclusions:
Whole blood can be used interchangeably with plasma for more convenient and less time and labor-consuming testing of hs-cTnI on the PATHFAST instrument.
More Related Videos
10:01Improvement of a Closed Chest Porcine Myocardial Infarction Model by Standardization of Tissue and Blood Sampling Procedures
Published on: March 12, 2018
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...