Pump position and thrombosis in ventricular assist devices: Correlation of radiographs and CT data

Philipp Aigner1,2, Thomas Schlöglhofer1,2,3, Lea Carmen Plunger1

  • 1Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria.

Insights

Malpositioning of left ventricular assist devices (LVAD) can cause pump thrombosis. Routine X-rays can identify HVAD pump malposition, predicting thrombosis risk in these patients.

Area of Science:

  • Cardiology
  • Medical Imaging
  • Biomedical Engineering

Background:

  • Left ventricular assist devices (LVADs) are crucial for heart failure management.
  • LVAD malpositioning is a known risk factor for pump thrombosis.
  • Identifying LVAD malposition from clinical imaging is challenging.

Purpose of the Study:

  • To identify imaging parameters from routine X-rays and CT scans that correlate with pump thrombosis in LVAD patients.
  • To determine if radiographs can reliably detect malpositioned LVADs and predict thrombosis risk.

Main Methods:

  • Retrospective analysis of imaging data (X-ray, CT) from 115 LVAD patients (HeartmateII and HVAD).
  • Comparison of imaging parameters between patients with pump thrombosis (PT) and a matched control group (NT).
  • Analysis of inflow cannula angles (α and β) in CT scans and correlation with radiographic findings.

Main Results:

  • In HVAD patients, specific radiographic measurements (pump body area, minor axis) and inflow cannula visibility predicted pump thrombosis.
  • CT analysis revealed significant differences in inflow cannula angle α between PT and NT groups in HVAD patients.
  • Routine radiographs demonstrated a correlation with CT angles, indicating their utility in identifying malpositioned LVADs.
  • No specific radiographic parameters were associated with pump thrombosis in HeartmateII patients.

Conclusions:

  • Routine radiographs can effectively identify malpositioned LVADs, particularly the HVAD.
  • Posterior rotation of the HVAD inflow cannula, detectable on radiographs, is associated with an increased risk of pump thrombosis.
  • Ensuring proper inflow cannula alignment is critical for preventing LVAD pump thrombosis.