In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
Pump position and thrombosis in ventricular assist devices: Correlation of radiographs and CT data
Philipp Aigner1,2, Thomas Schlöglhofer1,2,3, Lea Carmen Plunger1
1Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria.
Insights
Malpositioning of left ventricular assist devices (LVAD) can cause pump thrombosis. Routine X-rays can identify HVAD pump malposition, predicting thrombosis risk in these patients.
Area of Science:
- Cardiology
- Medical Imaging
- Biomedical Engineering
Background:
- Left ventricular assist devices (LVADs) are crucial for heart failure management.
- LVAD malpositioning is a known risk factor for pump thrombosis.
- Identifying LVAD malposition from clinical imaging is challenging.
Purpose of the Study:
- To identify imaging parameters from routine X-rays and CT scans that correlate with pump thrombosis in LVAD patients.
- To determine if radiographs can reliably detect malpositioned LVADs and predict thrombosis risk.
Main Methods:
- Retrospective analysis of imaging data (X-ray, CT) from 115 LVAD patients (HeartmateII and HVAD).
- Comparison of imaging parameters between patients with pump thrombosis (PT) and a matched control group (NT).
- Analysis of inflow cannula angles (α and β) in CT scans and correlation with radiographic findings.
Main Results:
- In HVAD patients, specific radiographic measurements (pump body area, minor axis) and inflow cannula visibility predicted pump thrombosis.
- CT analysis revealed significant differences in inflow cannula angle α between PT and NT groups in HVAD patients.
- Routine radiographs demonstrated a correlation with CT angles, indicating their utility in identifying malpositioned LVADs.
- No specific radiographic parameters were associated with pump thrombosis in HeartmateII patients.
Conclusions:
- Routine radiographs can effectively identify malpositioned LVADs, particularly the HVAD.
- Posterior rotation of the HVAD inflow cannula, detectable on radiographs, is associated with an increased risk of pump thrombosis.
- Ensuring proper inflow cannula alignment is critical for preventing LVAD pump thrombosis.
Abstract:
Malpositioning of left ventricular assist devices (LVAD) is a risk factor for thrombosis, but its identification from clinical imaging remains challenging. X-rays and CT scans were analyzed and parameters identified that correlated to pump thrombosis. Retrospective imaging data of patients (n = 115) with HeartmateII (HMII) or HVAD were analyzed in two groups (pump-thrombosis PT, n = 15 vs matched control group NT, n = 15) using routine X-rays and CT scans. In CT, directional deviations of the inflow cannula in three-chamber and two-chamber view (α and β angles) were identified. In HVAD PT frontal radiographs showed reduced pump body area and smaller minor axis (PT 41.3 ± 4.8 mm vs NT 34.9 ± 6.0 mm, p = 0.026), and in the lateral radiographs the visibility of the inflow cannula served as a predictive parameter for PT. In HMII patients, no parameters were associated with PT. The angle α differed significantly (NT -1.2 ± 7.5°, PT -22.0 ± 4.7°, p = 0.006) in HVAD patients. Further, correlations of x-ray parameters with CT angles α and β showed that radiographs can be used to identify malpositioned pumps. Well-aligned inflow cannula positions are essential. HVAD patients with a posterior rotation of the inflow cannula have a higher risk of pump thrombosis. This risk can reliably be identified from routine radiographs.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT