Near-infrared spectroscopy device selection affects intervention management for cerebral desaturation during
Kensuke Kobayashi1, Tadashi Kitamura2, Satoshi Kohira3
1Department of Cardiovascular Surgery, Kitasato University School of Medicine, 1-15-1 Kitasato, Minami, Sagamihara, Kanagawa, 252-0375, Japan. qzh01063@nifty.com.
Objective:
Currently, several near-infrared spectroscopy oximetry devices are used for detecting cerebral ischemia during cardiopulmonary bypass (CPB) surgery. We investigated whether two different models of near-infrared spectroscopy oximetry devices affect the assessment of cerebral ischemia and its management during CPB.
Methods:
From January 2017 to August 2017, 70 adult cardiovascular surgery cases were randomly assigned to 1 of 2 different near-infrared spectroscopy oximetry devices. The devices were INVOS 5100C (Medtronic, Minneapolis, MN, USA) (group I; n = 35) and FORE-SIGHT ELITE (CAS Medical Systems, Branford, CT, USA) (group F; n = 35).
Results:
There were no significant differences in patient characteristics. The rSO2 values were significantly higher for patients in group F than for patients in group I. Scalp-Cortex distance showed negative correlations with the mean rSO2 values in group I (P = 0.01). Interventions for low rSO2 during CPB for groups I and F were increase perfusion flow (13:5; P = 0.03), blood transfusion (7:1; P = 0.02), and both (6:1; P = 0.04), respectively. The Scalp-Cortex distance in group I was significantly longer in patients who required intervention than in patients who did not (17.1 ± 2.5 vs 15.1 ± 1.6 mm; P = 0.007).
Conclusions:
It is inappropriate to use the same intervention criteria for different near-infrared spectroscopy oximetry devices. Moreover, brain atrophy influence rSO2 values depending on device selection. It is important to note that inappropriate device selection may misguide perfusionists into performing unnecessary or excessive intervention during CPB.